CAT.NO LER0763
PRODUCT Rat BCL-3(B-Cell Leukemia/Lymphoma 3) ELISA Kit
DETECTION Sandwich ELISA, Double Antibody
PRICE KRW 480,000
  • Specifications
  • Manual
  • Contact
Product nameRat BCL-3 (B-Cell Leukemia/Lymphoma 3) ELISA Kit
Catalogue No.LER0763
Detection methodSandwich ELISA, Double Antibody
Size96T
Range6.25-400pg/ml
Sensitivity< 3.75pg/ml
Standard curve
Storage4 ¡É for 6 months
RecoveryMatrices listed below were spiked with certain level of BCL-3 and the recovery rates were calculated by comparing the measured value to the expected amount of BCL-3 in samples.
MatrixRecovery range(%)Average(%)
serum(n=5)85-9591
EDTA plasma(n=5)87-10395
heparin plasma(n=5)87-9892
LinearityThe linearity of the kit was assayed by testing samples spiked with appropriate concentration of BCL-3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample1:21:41:81:16
serum(n=5)89-102%85-94%85-97%86-104%
EDTA plasma(n=5)84-100%82-98%83-100%87-101%
heparin plasma(n=5)83-98%83-98%82-91%83-100%
CV(%)Intra-Assay: CV<8%
Inter-Assay: CV<10%
NoteFor Research Use Only